Precision and Accuracy in Measurements

Every meaningful chemical measurement carries two distinct questions inside it: how close is this value to the truth, and how close would it be if we measured again? Precision and accuracy are the twin answers to those questions, and confusing them is one of the most consequential errors a young chemist can make. A balance that reads 2.000 g every single time is precise, yet if its internal calibration is skewed, every one of those readings may sit far from the actual mass. Understanding this distinction transforms measurement from a mechanical act into an act of judgment.

This lesson builds directly on the rules of significant figures you have already studied. Significant figures tell you how many digits a result deserves; precision and accuracy tell you how much confidence those digits actually warrant. We will examine the classic student data set from the reference material, then extend the discussion into standard deviation, variance, systematic and random errors, calibration, bias, and the error bars that appear on scientific graphs. Along the way, examples from medical testing and elite sport will show that these ideas govern decisions far beyond the chemistry laboratory.

By the end of this lesson you should be able to look at any set of repeated measurements and say, with justification, whether the values cluster tightly, whether they centre on the true value, and what kind of error is most likely responsible for any discrepancy. That analytical habit is precisely what the next lesson on dimensional analysis and unit conversions will assume you possess, because converting units without understanding measurement uncertainty merely relocates error rather than removing it.

The Core Distinction Between Precision and Accuracy

Precision refers to the closeness of various measurements of the same quantity to one another. It is an internal property of the measurement process: it asks whether the instrument and the operator produce consistent results when the same thing is measured repeatedly. Accuracy, by contrast, is the agreement of a particular measured value with the true value of the result. Accuracy is an external property, because judging it requires knowledge of a reference truth that lies outside the measurement itself. A precise set of readings may be entirely inaccurate, and an accurate single reading may arise from a process so erratic that it cannot be trusted.

The reference data make this concrete. Suppose the true value for a result is 2.00 g. Student A reports 1.95 g and 1.93 g. These two values are close to each other, so they are precise, but both lie below the true value, so they are not accurate. Student B reports 1.94 g and 2.05 g; these are scattered widely and their average misses the mark, so they are neither precise nor accurate. Student C reports 2.01 g and 1.99 g, which cluster tightly around 2.00 g, making them both precise and accurate. The average of each student's pair, 1.940 g, 1.995 g and 2.000 g respectively, summarises the pattern neatly.

A scatter plot with a vertical reference line at 2.00 g; Student A's points at 1.95 and 1.93 g lie close together below the line, Student B's at 1.94 and 2.05 g are widely separated around it, showing precision without accuracy versus accuracy without precision.
Student A's two readings cluster tightly but sit below the 2.00 g true value, while Student B's readings straddle the truth but are far apart.Digital illustration created with Zartom Composer · Rahul Anand

Reading the Classic Data Set Carefully

Notice that Student B's average of 1.995 g is deceptively close to the true value of 2.00 g. If you judged only by the average, you might wrongly conclude that Student B performed well. The individual readings of 1.94 g and 2.05 g, however, differ by 0.11 g, which is a spread more than fifty times larger than the 0.002 g spread between Student C's readings. Averaging can hide poor precision, which is why a responsible analyst always reports the spread alongside the mean. A single number without an indication of its variability is an incomplete measurement.

This is also why the number of significant figures in a reported result must reflect the quality of the measurement, not merely the capacity of the display. A digital balance might show four decimal places, but if repeated readings wander in the second decimal place, the later digits are noise dressed as information. The significant-figure rules you learned previously tell you how to truncate a calculation; precision analysis tells you which digits were ever meaningful in the first place. The two skills work together, and neither is complete without the other.

A bar chart comparing the mean mass of Student A and Student B, both close to 2.00 g, with the spread of individual readings shown as the key difference the mean cannot reveal.
Both students average near 2.00 g, but the individual readings behind Student B's mean differ by 0.11 g, so the average alone conceals poor precision.Digital illustration created with Zartom Composer · Rahul Anand

Why the Distinction Matters in Practice

In routine laboratory work, precision is easy to monitor because it requires only repetition, while accuracy demands an independent standard. This asymmetry explains why sloppy procedures often produce beautifully consistent wrong answers. A burette with a worn stopcock may deliver slightly more volume than its scale indicates, yet every titration will agree closely with the last. The chemist sees excellent precision and feels confident, unaware that a systematic offset has corrupted every result in the notebook. Recognising this trap is the first step toward designing measurements that can actually detect their own flaws.

The practical remedy is to build accuracy checks into every procedure. Running a blank, analysing a certified reference material, or comparing against an independent method all provide the external anchor that repetition alone cannot supply. When such a check reveals a discrepancy, the analyst must decide whether the fault lies in the instrument, the reagents, the operator or the environment. That diagnostic reasoning, moving from a numerical symptom to a physical cause, is the professional core of analytical chemistry and the reason this lesson sits early in the course.

Worked Data

Precision and Accuracy Verdicts for Three Students

True value = 2.00 g; each student reports two measurements.

Student and Readings Average, Spread and Verdict
Student A: 1.95 g, 1.93 g Average 1.940 g; spread 0.02 g; precise but not accurate
Student B: 1.94 g, 2.05 g Average 1.995 g; spread 0.11 g; neither precise nor accurate
Student C: 2.01 g, 1.99 g Average 2.000 g; spread 0.02 g; both precise and accurate
Note:
  • Student B's average looks acceptable only because two large errors of opposite sign cancel.
  • Spread, not the average alone, is the honest indicator of precision.

Quantifying Precision with Standard Deviation and Variance

Describing precision as "the readings are close together" is qualitative and therefore insufficient for serious work. Statistics gives us two linked measures that convert the visual impression of clustering into a number. The variance is the average of the squared deviations of each reading from the mean, and the standard deviation is its square root. Squaring prevents positive and negative deviations from cancelling, while the square root returns the result to the original units, making it directly comparable with the measurements themselves. Together they form the standard language of precision in every physical science.

For a small set of ##n## repeated measurements ##x_1, x_2, \ldots, x_n## with mean ##\bar{x}##, the sample standard deviation is defined as shown below. The denominator ##n-1## rather than ##n## reflects the fact that the mean itself was estimated from the same data, consuming one degree of freedom. For the two-reading sets in our table, this formula gives Student A a standard deviation of about 0.014 g, Student B about 0.078 g, and Student C about 0.014 g, confirming numerically what the eye already suspected.

A plotted curve illustrating how the sample standard deviation grows with the spread of deviations from the mean, with the n minus 1 denominator noted.
The sample standard deviation divides the summed squared deviations by n minus 1, then takes the square root.Digital illustration created with Zartom Composer · Rahul Anand
###s = \sqrt{\dfrac{\sum_{i=1}^{n}(x_i - \bar{x})^2}{n-1}}###

Interpreting the Numbers Sensibly

A small standard deviation signals high precision, but it says nothing about accuracy unless the mean is also compared with a reference value. The standard deviation is therefore best reported as a companion to the mean, in the form ##\bar{x} \pm s##, which communicates both the central estimate and its reliability. Reporting 2.000 ± 0.014 g tells a reader far more than reporting 2.000 g alone. In professional practice, the uncertainty is often quoted at a stated confidence level, commonly one standard deviation or a 95% interval, and the choice must always be made explicit.

Variance is less intuitive because its units are squared, but it has a valuable property: variances of independent sources of error add directly. If a measurement is affected by both instrument noise and sample heterogeneity, the total variance is the sum of the individual variances. This additivity underpins uncertainty budgets in analytical laboratories and explains why reducing the largest single source of variance yields the greatest improvement in overall precision. A chemist who understands this can prioritise effort intelligently rather than polishing negligible contributions.

From Two Readings to Many

Two measurements are the minimum needed to see any spread at all, but they give a very unstable estimate of standard deviation. Professional protocols typically require five to ten replicates before quoting an uncertainty, because the standard deviation of a small sample fluctuates substantially from one experiment to the next. With only two readings, a single aberrant value can double the apparent uncertainty. Students should therefore treat the two-reading examples in the table as illustrations of principle rather than as models of good experimental design.

When replicate numbers grow large, the standard error of the mean becomes the relevant quantity for comparing averages. It equals the standard deviation divided by the square root of ##n##, so quadrupling the number of replicates halves the uncertainty on the mean. This square-root relationship is one of the most useful facts in experimental science: it tells you that improving precision by repetition has diminishing returns, and that beyond a certain point it is wiser to improve the instrument or the method instead.

Systematic and Random Errors: The Two Sources of Imperfection

Every deviation from truth can be traced to one of two broad categories of error. Random errors fluctuate unpredictably in both magnitude and sign from one measurement to the next, arising from vibration, temperature drift, electrical noise, or the finite resolution of a reading scale. Systematic errors, by contrast, push results consistently in one direction, arising from miscalibrated instruments, impure reagents, faulty technique, or an incorrect assumption built into the method. The two behave so differently that identifying which dominates is the essential diagnostic step in any measurement problem.

The consequences for precision and accuracy are strikingly different. Random errors degrade precision, producing scatter around the mean, but they tend to average out when many readings are taken, so the mean can still be accurate. Systematic errors leave precision untouched, since every reading is displaced by the same amount, but they corrupt accuracy in a way that repetition cannot repair. This is why Student A's tight pair of low readings points to a systematic error, while Student B's scattered pair points to random error. The pattern of the data reveals the character of the culprit.

A scatter plot with a reference line at the true value; one group of points is tightly clustered but offset above the line, another is centred on the line but widely scattered, illustrating systematic versus random error.
Random error scatters readings around the true value, while systematic error shifts the whole cluster away from it.Digital illustration created with Zartom Composer · Rahul Anand

Diagnosing the Error Type from Data

A useful diagnostic rule is to compare the spread with the offset. If the readings cluster tightly but their mean sits consistently away from the reference value, suspect a systematic error and investigate calibration, reagent purity or procedural assumptions. If the readings scatter widely but their mean lands near the reference, suspect random error and consider more replicates, better shielding or a more stable environment. If both symptoms appear, as with Student B, the method needs attention on two fronts simultaneously, and averaging alone will not rescue it.

Real instruments often show a mixture. A pH meter may drift slowly during a session, which is systematic within a short window but random across a long one, while its electrode also produces rapid flicker that is purely random. Skilled analysts learn to separate these timescales, recalibrating between batches to control the slow drift while averaging several readings to suppress the fast flicker. This layered approach to error control is what distinguishes a competent measurement scientist from someone merely following a recipe.

Common Mistakes Worth Anticipating

The most frequent student error is to assume that repeating a measurement improves accuracy. Repetition improves precision and reduces the influence of random error, but a systematic offset survives every repetition intact. A second common mistake is to treat a single outlier as a mistake and discard it without justification; outliers may be genuine evidence of an intermittent fault or a contaminated sample. A third is to confuse the resolution of a display with the uncertainty of a measurement, forgetting that the whole apparatus contributes to the total error.

Another subtle error is to average readings taken under changing conditions, such as a balance that warms up during the session. The mean of such a drifting series is not a meaningful estimate of anything, because the underlying quantity being measured has effectively changed. Good practice requires that conditions be stable before replicates are collected, and that any drift be documented rather than silently averaged away. Careful record-keeping of conditions is therefore as much a part of measurement as the reading itself.

Error Taxonomy

Random Versus Systematic Error at a Glance

How each error family affects precision, accuracy and remedy.

Feature Random Error and Systematic Error
Typical origin Noise, vibration, drift versus miscalibration, impure reagents, faulty method
Effect on precision Degrades precision versus leaves precision unaffected
Effect on accuracy Averages out with replicates versus persists in every replicate
Standard remedy More replicates, shielding, stable conditions versus calibration, blanks, reference materials
Note:
  • Repetition is a cure for random error only; it cannot remove a systematic offset.
  • Most real measurements contain both error types in different proportions.

Calibration, Bias and the Pursuit of Accuracy

Calibration is the process of comparing an instrument's readings against a known standard and adjusting it so that its output aligns with the truth. A balance is calibrated with certified masses, a pH meter with buffer solutions of known pH, and a spectrophotometer with standards of known absorbance. Calibration directly attacks systematic error, which is why it is the single most powerful tool for improving accuracy. Without calibration, an instrument's precision is merely a measure of how consistently it repeats its own mistakes.

Bias is the quantitative expression of systematic error: the consistent deviation of measured values from the true value, often expressed as the difference between the mean of many readings and the reference. A positive bias means the instrument reads high; a negative bias means it reads low. Bias can arise from the instrument itself, from the operator, from the sample matrix, or from the model used to interpret the signal. Because bias is directional, it can be measured, corrected and monitored over time, unlike random error, which can only be reduced.

How Calibration Actually Improves Accuracy

A single-point calibration adjusts the instrument at one reference value, correcting a simple offset but leaving any scale error untouched. A two-point calibration, using standards near the low and high ends of the working range, corrects both offset and slope, and is the minimum for serious quantitative work. Multi-point calibration with a fitted curve extends this to non-linear responses, such as those of ion-selective electrodes. The choice of calibration strategy should match the expected behaviour of the instrument and the accuracy required by the application.

A line plot comparing an uncorrected response with a one-point-calibrated response that is shifted but still wrongly sloped, and a two-point-calibrated response that coincides with the ideal identity line.
A one-point calibration corrects only the offset, while a two-point calibration corrects both offset and slope so the response matches the ideal line.Digital illustration created with Zartom Composer · Rahul Anand

Calibration is not a one-time ritual. Instruments drift as components age, electrodes foul, and lamps dim, so calibration must be repeated at defined intervals and verified with independent check standards between full calibrations. A check standard that falls outside its acceptance limits signals that the instrument has drifted and that recent results may need re-examination. This regime of periodic verification is what allows a laboratory to claim traceability, the documented chain linking its measurements back to national or international standards.

Bias in the Broader Sense

Bias is not confined to hardware. Sampling bias occurs when the portion of material analysed does not represent the whole, as when a heterogeneous ore is sampled only from its surface. Observer bias occurs when an analyst unconsciously rounds readings toward an expected value, a well-documented phenomenon in manual timing and reading of analogue scales. Method bias occurs when the chemistry of the analysis systematically favours one species over another. Each of these requires a different countermeasure, from randomised sampling to blind reading protocols to independent cross-method validation.

Recognising bias is often harder than correcting it, because a biased measurement system looks perfectly well-behaved from the inside. The only reliable defence is external comparison: analysing a certified reference material, participating in inter-laboratory proficiency schemes, or checking against a fundamentally different technique. When two independent methods agree, confidence in accuracy rises sharply; when they disagree, the investigation that follows usually exposes a bias that neither method could have detected alone.

Precision and Accuracy Beyond the Laboratory

The vocabulary of this lesson travels well beyond chemistry. In medical testing, a diagnostic assay must be both precise, giving the same result on repeated runs of the same sample, and accurate, agreeing with the patient's true physiological state. A test that is precise but biased will consistently misclassify patients, while a test that is accurate on average but imprecise will produce erratic results that erode clinical trust. Regulatory bodies therefore demand formal studies of both repeatability and trueness before any assay enters routine use.

Sport offers equally vivid illustrations. A timing system at a athletics meeting may be precise to a thousandth of a second yet be systematically offset if its start signal is delayed, producing beautifully consistent but uniformly wrong times. A marksman's shots may cluster tightly in the wrong part of the target, the classic picture of precise but inaccurate performance, correctable only by adjusting the sights rather than by shooting more carefully. In both settings, the distinction between scatter and offset determines whether the remedy lies in technique or in equipment.

Lessons from Medical Testing

Consider a laboratory measuring blood glucose. If the analyser repeatedly returns 5.4 mmol/L for a sample whose true concentration is 5.0 mmol/L, the assay is precise but biased high by 0.4 mmol/L. Every patient measured on that analyser will appear slightly more hyperglycaemic than they are, and at the diagnostic threshold this bias could push borderline individuals into a treatment category. The remedy is recalibration against a reference method, followed by verification that the bias has been removed across the clinically relevant range.

A bar chart comparing the true blood glucose concentration of 5.0 mmol/L with the analyser's repeated reading of 5.4 mmol/L, showing a consistent positive bias.
An analyser returning 5.4 mmol/L for a true 5.0 mmol/L sample is precise but biased high by 0.4 mmol/L.Digital illustration created with Zartom Composer · Rahul Anand

Conversely, an assay with a standard deviation of 0.5 mmol/L but no bias will occasionally misclassify a borderline patient in either direction, but the errors will not systematically favour one diagnosis. Clinicians and regulators weigh these two failure modes differently depending on the consequences of misclassification, which is why analytical performance goals are set separately for bias and imprecision. The statistical machinery of this lesson thus directly informs patient safety and public health policy.

Lessons from Sport and Engineering

In engineering, the same logic governs tolerances. A machined component may be produced with excellent repeatability, yet if the machine tool is offset, every part will be uniformly out of specification. Quality control therefore monitors both the spread of a production run, using control charts, and its centring relative to the nominal dimension. A process that is precise but off-centre is easier to fix than one that is centred but wildly variable, because a single adjustment to the machine can recentre the entire distribution.

This asymmetry between correcting bias and reducing scatter is a recurring theme. Bias correction is often a matter of a single calibration or adjustment, while scatter reduction may require redesigning the entire process. Understanding which problem you face therefore determines whether the solution is cheap or expensive, quick or slow. The ability to read a data set and identify the dominant failure mode is a transferable skill of considerable commercial and scientific value.

Applied Contexts

Precision and Accuracy Across Four Domains

The same two questions recur wherever numbers inform decisions.

Domain Precision Concern and Accuracy Concern
Clinical assay Repeatability across runs versus agreement with reference method
Athletics timing Consistency of the clock versus correct start and stop signals
Manufacturing Spread of a production run versus centring on nominal dimension
Environmental monitoring Sensor noise versus calibration drift over deployment
Note:
  • In every domain, bias correction is usually cheaper than scatter reduction.
  • Regulatory standards typically specify separate limits for bias and imprecision.

Visualising Uncertainty with Error Bars

Numbers in a table can conceal the very variability that matters most, which is why scientists plot data with error bars. An error bar is a graphical mark extending above and below a plotted point to indicate the uncertainty in that value. The length of the bar may represent a standard deviation, a standard error, a confidence interval or a manufacturer's tolerance, and the caption must always state which. Without that statement, an error bar is decorative rather than informative, and readers may draw unjustified conclusions about significance.

Error bars transform a graph from a set of points into a statement about reliability. Two means whose error bars overlap substantially are unlikely to differ significantly, whereas two means separated by several bar lengths almost certainly do. This visual heuristic is not a substitute for a formal statistical test, but it communicates the essential comparison at a glance. Learning to read error bars critically is therefore as important as learning to compute them, because graphs persuade far more powerfully than tables.

Choosing What the Bar Represents

The standard deviation bar describes the spread of individual measurements, while the standard error bar describes the uncertainty of the mean and is smaller by a factor of ##\sqrt{n}##. Confidence interval bars are wider still, typically spanning about twice the standard error for a 95% interval with moderate sample sizes. Because these three quantities differ by known factors, a graph can look dramatically more or less convincing depending on which is chosen. Honest reporting requires the choice to be stated and justified rather than selected to flatter the result.

A bar chart comparing the lengths of standard deviation, standard error and confidence interval error bars for identical data, showing how the choice of bar changes the apparent uncertainty.
For the same data the standard deviation bar is widest, the standard error bar is smaller by the square root of n, and the confidence interval bar is wider again.Digital illustration created with Zartom Composer · Rahul Anand

Error bars can also be asymmetric, particularly when data are transformed logarithmically or when uncertainties are multiplicative. In such cases the upper and lower extents differ, and plotting software must be instructed accordingly. Ignoring asymmetry can misrepresent the data, especially near detection limits where the uncertainty in one direction may be unbounded. Careful attention to these details separates a publication-quality figure from a misleading one.

From Error Bars to Scientific Judgement

When error bars are properly understood, they change how results are interpreted. A difference between two treatments that looks impressive in a bar chart may shrink to insignificance once the uncertainty is displayed, while a modest-looking difference with tiny error bars may be highly reliable. The eye is drawn to the height of bars, but the scientist must be drawn to the overlap of their uncertainty ranges. This habit of looking at uncertainty first, and central values second, is a hallmark of mature quantitative reasoning.

The same discipline applies to the worked examples in this course. When you report a titration volume or a calculated molar mass, you should be able to state the uncertainty attached to it and to sketch, mentally at least, the error bar that would accompany it on a graph. That reflex will serve you well in the next lesson, where unit conversions must be performed without silently inflating or deflating the precision of the final answer.

Graph Literacy

What Different Error Bars Actually Mean

The same data can look very different depending on the chosen statistic.

Bar Type Meaning and Relative Length
Standard deviation Spread of individual readings; longest of the three
Standard error of mean Uncertainty of the mean; smaller by factor of square root of n
95% confidence interval Range likely to contain the true mean; roughly twice the standard error
Tolerance or specification Manufacturer or regulatory limit; not a statistical estimate
Note:
  • Always state in the caption which statistic the error bar represents.
  • Overlapping bars suggest no significant difference but do not prove it.

Bringing It Together: A Practical Framework for Judging Measurements

We can now assemble the lesson into a working procedure. First, collect replicates under stable conditions and compute the mean and standard deviation, reporting both. Second, compare the mean with a reference value obtained from a certified standard or an independent method, and compute the bias. Third, inspect the pattern: tight scatter with a consistent offset indicates systematic error, while wide scatter around a correct mean indicates random error. Fourth, apply the appropriate remedy, calibrating for bias or increasing replicates and controlling conditions for scatter. Finally, report the result with its uncertainty and, where useful, plot it with clearly labelled error bars.

This framework is deliberately simple, but it captures the essential logic that separates a measurement from a mere number. It also prepares you for the next lesson, because unit conversions performed on a well-characterised measurement preserve its meaning, whereas conversions performed on an unexamined number merely propagate ignorance. The habit of asking both questions, how close together and how close to truth, should accompany every quantitative result you produce from this point onward in the course.

Worked Illustration with a Full Data Set

Suppose five replicate titrations give volumes of 24.82, 24.86, 24.79, 24.84 and 24.81 mL. The mean is 24.824 mL and the sample standard deviation is approximately 0.027 mL, indicating good precision. If a certified standard predicts 24.90 mL, the bias is about 0.076 mL, roughly three standard deviations, which is unlikely to be chance. The correct conclusion is that the method is precise but biased low, and the remedy is to investigate the burette calibration or the indicator endpoint rather than to collect more replicates.

A scatter plot of five titration volumes between 24.79 and 24.86 mL with a reference line at 24.90 mL, showing a tight cluster offset below the certified value.
Five titration volumes cluster tightly near 24.82 mL but sit about 0.08 mL below the certified 24.90 mL, indicating good precision with a small systematic bias.Digital illustration created with Zartom Composer · Rahul Anand

Had the same five readings scattered between 24.60 and 25.05 mL while averaging 24.90 mL, the diagnosis would reverse: acceptable accuracy with poor precision, calling for better technique, degassing of solutions or temperature control. The contrast between these two hypothetical data sets, identical in mean but opposite in character, is the heart of this lesson. A chemist who can distinguish them at a glance possesses a skill that no amount of formula memorisation can replace.

Preparing for Dimensional Analysis

The next lesson will multiply measurements by unit factors to convert between systems. Every such conversion carries the precision of its inputs forward, and the significant-figure rules you have learned govern how the result is rounded. But rounding is only meaningful if the underlying measurement was characterised in the first place. A conversion applied to a biased value produces a biased value in new units, and a conversion applied to a noisy value produces noise in new units. Precision and accuracy analysis is therefore the foundation on which all subsequent quantitative work rests.

As you proceed, keep the two questions in mind whenever a number appears: how reproducible is it, and how close to truth is it likely to be? Ask also what kind of error dominates and what evidence would reveal it. These questions cost nothing to ask and frequently save an entire experiment from being wasted. They are, in the truest sense, the habits of a scientist rather than the habits of a calculator.

Decision Guide

Diagnostic Framework for Measurement Problems

Match the observed pattern to the likely cause and remedy.

Observed Pattern Likely Cause and Remedy
Tight cluster, offset mean Systematic error; recalibrate and check reagents
Wide scatter, correct mean Random error; more replicates and stable conditions
Wide scatter, offset mean Both error types; method overhaul required
Tight cluster, correct mean Both precise and accurate; document and proceed
Note:
  • Always verify the reference value before declaring a bias.
  • Document conditions alongside readings so drift can be diagnosed later.
Formula Sheet

Key Formulas and Their Roles

The statistical tools introduced in this lesson.

Quantity Expression and Interpretation
Mean Sum of readings divided by n; central estimate
Variance Mean squared deviation; additive across independent sources
Standard deviation Square root of variance; measures precision in original units
Standard error Standard deviation divided by square root of n; uncertainty of mean
Bias Mean minus reference value; measures accuracy
Note:
  • Use n minus one in the denominator for sample standard deviation.
  • Report mean and standard deviation together for a complete picture.

For learners who wish to automate these calculations, the short Python routine below reproduces the mean, sample standard deviation and bias for the titration example discussed above. Running it on your own data is an excellent way to internalise the relationship between the formulas and the numbers they produce.


import statistics

readings = [24.82, 24.86, 24.79, 24.84, 24.81]
reference = 24.90

mean = statistics.mean(readings)
sd = statistics.stdev(readings)
bias = mean - reference

print(f"Mean = {mean:.3f} mL")
print(f"Standard deviation = {sd:.3f} mL")
print(f"Bias = {bias:+.3f} mL")